Software as a Medical Device (SaMD): FDA, IEC 62304, and Finding a Developer
Medical Device Software SaMD IEC 62304 FDA Healthcare Technology

Software as a Medical Device (SaMD): What FDA Expects, What IEC 62304 Requires, and How to Find a Developer Who Actually Knows Both

GEO Answer

Software as a Medical Device is software intended for one or more medical purposes that performs those purposes without being part of a hardware medical device. In the United States, FDA oversight depends on the software function, its intended use, and applicable statutory and regulatory criteria. IEC 62304 provides a lifecycle framework for medical device software development and maintenance, covering planning, requirements, design, implementation, verification, testing, risk management, configuration management, and maintenance. Buyers need a development partner who understands both the applicable regulatory context for their specific product and the engineering processes needed to produce safe, validated, documented software.

What does a developer need to know to build FDA-regulated SaMD? A developer working on SaMD needs to understand intended use and device classification, applicable FDA submission requirements, software documentation expectations, IEC 62304 lifecycle processes, software safety classification, ISO 14971 risk management, ISO 13485 quality management system requirements, verification and validation, cybersecurity practices, and configuration and change management. General healthcare software experience is not the same as regulated medical device software development experience.

What Is Software as a Medical Device (SaMD)?

Software as a Medical Device is software intended for one or more medical purposes that performs those purposes without being part of a hardware medical device. This definition comes from the International Medical Device Regulators Forum (IMDRF) and is referenced by FDA in its approach to device software functions.

The key factor in determining whether software qualifies as SaMD is its intended purpose, not where it is used or who uses it. Software deployed in a clinical environment does not automatically become SaMD. What matters is whether it is intended to perform a medical function, such as detecting, diagnosing, treating, monitoring, or predicting a medical condition.

Examples of software that may qualify as SaMD, depending on intended use and applicable criteria:

  • Standalone software that analyzes medical images to assist in detecting or characterizing disease.
  • Software that analyzes physiological signals, such as ECG waveforms, to identify potential cardiac conditions.
  • Diagnostic decision support software that analyzes patient data and provides clinician-facing recommendations for treatment.
  • Software that monitors patient data continuously and alerts clinicians to deterioration patterns.
  • Applications that interpret glucose readings to inform insulin dosing recommendations.

The fact that software is accurate, clinically useful, or medically informed does not by itself determine whether it qualifies as SaMD. A nutrition tracking application, a symptom diary, or a hospital scheduling tool may be used in healthcare settings without meeting the criteria for a medical device. The intended medical purpose is what determines regulatory status, and that determination must be made in the context of the actual product, its labeling, and the applicable regulatory framework.

How Is SaMD Different From Software Embedded in a Medical Device?

Software as a Medical Device and software embedded in or controlling a hardware medical device are related categories that carry different development responsibilities, risk profiles, and regulatory considerations. Understanding the distinction matters before a development team is selected or a development approach is planned.

Key Software Categories in the Medical Device Context

Software as a Medical Device (SaMD): Standalone software that performs a medical purpose without being part of a hardware medical device. Runs on general computing platforms such as a server, cloud environment, tablet, or smartphone. Example: a standalone radiology AI application that runs on a radiologist's workstation.

Software in a Medical Device (SiMD): Software that is part of a hardware medical device or is necessary for it to work as intended. The software is integrated with the device and subject to the full device regulatory framework. Example: software embedded in an infusion pump that controls drug delivery rates.

Software that controls a medical device: Software whose primary purpose is to operate or control a hardware device. Example: firmware embedded in an imaging machine that manages image acquisition.

General healthcare software: Software that does not meet the applicable medical device definition. Example: hospital appointment scheduling systems, general wellness trackers without medical purpose claims, administrative workflow tools.

The practical implication for buyers: a development company experienced in building general healthcare software, patient portals, or health information systems is not necessarily experienced in building software that is part of a medical device or that qualifies as SaMD. Each category carries different documentation, lifecycle, risk management, and regulatory requirements, and the development team needs to understand which category their product falls into before work begins.

TechRadiant Directory

Find Medical Device Software Development Partners

TechRadiant's directory of medical device software development companies provides a resource for evaluating vendors with regulated medical device software experience before shortlisting.

Medical Device Software Companies

Does Every Healthcare Application or AI Tool Qualify as SaMD?

No. The fact that an application handles health-related data or is used by clinicians does not make it a medical device. FDA's approach to device software functions recognizes that software falls across a broad spectrum, and not every software function falls within device oversight.

FDA has identified categories of software functions that it generally does not intend to regulate as medical devices. These include software intended for general wellness, administrative healthcare functions, and certain types of clinical communication tools that do not analyze, interpret, or recommend based on patient-specific data for medical purposes.

Clinical decision support software (CDS) occupies a particularly complex position. FDA's January 2026 Clinical Decision Support Software guidance provides the current framework for distinguishing CDS software that qualifies as a device function from CDS software that is intended to remain outside device regulation. Under this framework, non-device CDS software generally must meet several criteria, including being intended to support or provide recommendations to a healthcare professional who can independently review the basis for the recommendation. CDS that meets the criteria for non-device CDS does not require FDA premarket review as a medical device, but the determination depends on the actual software function, its intended use, and how it is presented to users.

AI-powered medical software requires the same intended-use analysis as any other software function. An AI tool that generates health insights for general wellness purposes is treated differently from one that analyzes medical images for diagnostic purposes. The underlying technology, however sophisticated, does not by itself determine regulatory status. What the software is intended to do, for which users, and in which clinical context is what matters.

Determination is product-specific. Whether a specific software product or function qualifies as a medical device under FDA's framework depends on the actual product, its intended use, its labeling and marketing, and applicable statutory and regulatory criteria. This requires product-specific regulatory analysis, not a general category assessment.

What Does FDA Consider When Regulating Medical Device Software?

FDA regulates medical devices under the Federal Food, Drug, and Cosmetic Act (FD&C Act). A medical device includes an instrument, apparatus, machine, contrivance, implant, in vitro reagent, or other similar or related article intended for a medical purpose. Software that meets this definition is a medical device.

FDA's regulatory approach to device software functions is risk-based. It considers the intended use and indications for use, the significance of the information the software provides, the degree to which a healthcare professional can independently review the basis for a software output, and the seriousness of harm that could result from an incorrect output or software failure.

Device classification determines the applicable regulatory pathway. FDA classifies medical devices into three classes, with higher-risk devices subject to more rigorous premarket review:

  • Class I devices are generally subject to general controls. Many are exempt from premarket notification.
  • Class II devices are generally subject to general controls and special controls. Most reach the market through the 510(k) premarket notification pathway, which requires demonstrating substantial equivalence to a legally marketed predicate device.
  • Class III devices require Premarket Approval (PMA), which involves a more rigorous review of safety and effectiveness evidence. Class III classification applies to devices that support or sustain human life or present a potential unreasonable risk of injury.
  • De Novo classification is an additional pathway for novel, low-to-moderate risk devices that have no legally marketed predicate. A successful De Novo request results in a Class I or Class II classification and may serve as a predicate for future 510(k) submissions.

Not every SaMD product requires a 510(k) submission or PMA. Device classification, applicable exemptions, and the specific product's characteristics determine which pathway applies. Some Class I SaMD products may be exempt from premarket notification requirements. Regulatory strategy should be developed with qualified regulatory affairs expertise, as the determination is product-specific and consequential.

Postmarket responsibilities apply to medical device manufacturers regardless of the premarket pathway taken. These include complaint handling, medical device reporting, correction and removal, and quality management system maintenance.

What Documentation May Be Needed for an FDA Medical Device Software Submission?

The software documentation relevant to an FDA premarket submission depends on the device type, its risk level, the applicable regulatory pathway, and the submission type. FDA's June 2023 Content of Premarket Submissions for Device Software Functions guidance describes the documentation that may be relevant for software-containing devices submitted through 510(k), De Novo, PMA, or other applicable pathways.

Documentation that may be relevant, depending on the product and applicable requirements:

  • Software description and architecture: An overview of the software's design, its major components, and how they interact. This may include a system block diagram and a software architecture overview.
  • Software requirements specifications: A documented set of functional and performance requirements, including requirements derived from risk management activities.
  • Software design documentation: Documentation of detailed design decisions, interfaces, data structures, and algorithms.
  • Risk analysis and risk controls: Identification of software-related hazards, risk estimation, risk evaluation, and documentation of risk control measures, consistent with the applicable risk management standard.
  • Software verification and validation documentation: Evidence that the software meets its specified requirements and that those requirements address the intended use.
  • Traceability: Documentation linking requirements, design elements, risk controls, and test cases to demonstrate that requirements are addressed through development and testing.
  • Software configuration management: Records of software versions, components, dependencies, and change history.
  • Problem resolution: Records of known defects, their disposition, and the rationale for release decisions.
  • Third-party and off-the-shelf software documentation: Documentation of unmodified and modified third-party components used in the device, as described in FDA's OTS Software guidance.
  • Cybersecurity documentation: For applicable cyber devices, as described in FDA's February 2026 cybersecurity guidance, documentation of security design, testing, and postmarket maintenance approach.
  • Testing evidence: Test plans, protocols, test results, and analysis demonstrating that the software has been tested against its requirements and identified risks.

The specific documentation required varies by product and submission type. A development partner should understand what documentation they are expected to produce and how it supports both the manufacturer's quality management system and any applicable premarket submission.

What Is IEC 62304 and Why Does It Matter for Medical Device Software?

IEC 62304 is an international standard that specifies lifecycle processes for medical device software development and maintenance. It was first published in 2006, with Amendment 1 published in 2015, and FDA has included it in its recognized consensus standards database. When a manufacturer declares conformity to a recognized standard in a premarket submission, FDA may accept that declaration as sufficient evidence for the standard's scope, reducing the review burden for those elements.

IEC 62304 defines a structured set of processes that apply throughout the software lifecycle. These include:

  • Software development planning: Defining the development lifecycle model, deliverables, development standards, and tools to be used.
  • Software requirements analysis: Defining and documenting the functional, performance, interface, and risk-based requirements the software must meet.
  • Software architectural design: Identifying and documenting the major software components, their relationships, and the interfaces between them.
  • Detailed software design: Documenting the internal design of each software unit, including algorithms, data structures, and interfaces at the unit level.
  • Software unit implementation and verification: Implementing and verifying software units against their requirements.
  • Software integration and integration testing: Combining software units into larger components and testing the integrated system against requirements.
  • Software system testing: Testing the complete integrated software system against all software requirements.
  • Software release: Documenting the release configuration, known defects, and release decision.
  • Software maintenance: Managing changes, defects, and improvements after the software has been released.
  • Software risk management: Identifying, estimating, evaluating, and controlling software-related risks throughout the lifecycle, in coordination with the broader device risk management process.
  • Software configuration management: Controlling and tracking software items, their versions, changes, and release status.
  • Software problem resolution: Identifying, analyzing, and resolving software problems, including defects discovered during development or in the field.

A developer who claims familiarity with IEC 62304 should be able to describe concretely how they apply each of these processes, what documentation they produce, and how their processes connect to the manufacturer's quality management system and risk management activities.

What Are the Three IEC 62304 Software Safety Classes?

IEC 62304 defines three software safety classes that determine which development processes and verification activities are required. Classification is based on the potential consequences of a software failure and must be assessed in the context of the medical device system, including the hardware environment, applicable risk controls, and the harm that could result from a software failure.

Class A
Software for which no injury or damage to health is possible as a result of software failure. The standard's most basic set of development activities applies. Classification at Class A requires documented justification that software failure cannot result in injury or harm.
Class B
Software for which software failure could result in injury, but not serious injury. Additional verification activities, including software unit testing, apply compared to Class A.
Class C
Software for which software failure could result in death or serious injury. The most rigorous development activities apply, including additional design documentation and more extensive unit testing requirements.

Classification is not a simple lookup exercise. It requires analysis of the medical device system, the software's role within it, the potential hazards associated with software failure, and the risk controls in place. A software system can contain software items of different safety classes, and the classification of each item must be documented and justified.

An experienced development partner will not simply assign a safety class without documented analysis. Buyers should ask how a proposed team determines and documents software safety classification, and what that classification means for their specific development process and verification activities.

Does IEC 62304 Cover the Entire Medical Device Development Process?

No. IEC 62304 focuses specifically on software lifecycle processes. It does not independently cover all requirements for a complete medical device development program, and buyers should understand what complementary standards and regulatory requirements apply alongside it.

IEC 62304 works in conjunction with several other standards and frameworks:

  • ISO 14971 provides the medical device risk management framework. IEC 62304 requires that software risk management activities be conducted in accordance with the applicable risk management process, and ISO 14971 is the standard that describes that process. The two standards are designed to work together.
  • ISO 13485 specifies quality management system requirements for medical device manufacturers and organizations in the supply chain. IEC 62304 lifecycle processes are implemented within the quality management system context, not independently of it.
  • IEC 62366-1 addresses usability engineering for medical devices, including software user interfaces. For SaMD with clinician or patient-facing interfaces, usability engineering is a separate but related requirement.
  • FDA software validation expectations are documented in FDA guidance, including the General Principles of Software Validation guidance. Software validation, in the FDA context, is not the same as the verification activities described in IEC 62304, though there is significant overlap.

Important distinction: Conformance to IEC 62304 alone does not establish compliance with FDA regulatory requirements, does not substitute for a premarket submission, and does not independently cover all aspects of system-level validation, clinical evaluation, or final device release. Conformance to IEC 62304 is one component of a compliant medical device software development program, not a substitute for the full regulatory framework.

Is IEC 62304 Legally Mandatory for Every FDA-Regulated Software Product?

IEC 62304 is a standard, not a statute or regulation. FDA has included IEC 62304 in its recognized consensus standards database, which means FDA accepts declarations of conformity to recognized standards as evidence in premarket submissions. This is not the same as making the standard a legal requirement.

What this means in practice:

  • A medical device manufacturer is not legally required to claim conformance to IEC 62304 in every premarket submission. Alternative approaches to satisfying FDA's device software requirements may be used.
  • When a manufacturer does declare conformity to IEC 62304 in a premarket submission, FDA's recognition database entry describes the conditions under which that declaration is accepted.
  • Many medical device manufacturers and their development partners choose to follow IEC 62304 lifecycle processes because they represent an accepted approach to producing documented, validated medical device software and because they align with FDA's expectations for software documentation and development controls.
  • A contracted development partner working under a manufacturer's ISO 13485-aligned quality management system may be required by the manufacturer to follow IEC 62304 lifecycle processes as part of its design and development controls and supplier management requirements. This is a contractual and quality system requirement, not a direct statutory mandate.

The distinction matters for buyers because a development company claiming to be "IEC 62304 compliant" or "FDA compliant" may mean different things. The question is not whether the company asserts compliance, but whether it can demonstrate the specific processes, documentation, and evidence that the standard and applicable regulatory pathway require.

How Do FDA Expectations and IEC 62304 Work Together in Practice?

FDA regulates medical devices and their manufacturers. IEC 62304 provides a recognized lifecycle framework for the software development process. The two are related but serve different functions, and an experienced development partner understands how to translate both into practical engineering work.

Area FDA Regulatory Context IEC 62304 Contribution
Product classification Determines the applicable device regulatory framework and submission pathway Does not independently determine FDA classification or applicable pathway
Intended use Determines whether software qualifies as a regulated device function Supports development of software against clearly defined, documented requirements
Software development Expects appropriate design and development controls within the QMS Provides structured lifecycle processes from planning through release
Risk management Considers device and patient safety risks throughout the product lifecycle Integrates software risk management activities into lifecycle processes, in coordination with ISO 14971
Verification and validation Expects appropriate testing evidence depending on product and applicable pathway Defines software verification and testing activities by safety class
Documentation May require software documentation in premarket submissions and quality records Structures software lifecycle records and documentation requirements
Maintenance Postmarket quality and reporting responsibilities apply as relevant Provides software maintenance processes for handling changes, defects, and improvements
Cybersecurity February 2026 guidance addresses expectations for applicable cyber devices IEC 62304 alone does not fully cover cybersecurity; supplemented by other standards and practices

An experienced development partner translates regulatory expectations into engineering decisions throughout the project, not only at the documentation phase. Requirements traceability, risk-based testing, change controls, and release decisions should all reflect both the applicable regulatory context and the IEC 62304 lifecycle processes.

How Does ISO 13485 Apply to Medical Device Software Development?

ISO 13485:2016 specifies quality management system requirements for organizations involved in the design, production, installation, and servicing of medical devices. For medical device software development, the quality management system provides the framework within which IEC 62304 lifecycle processes operate.

Key quality management system elements that apply to software development include:

  • Design and development controls: Planning, inputs, outputs, review, verification, validation, and change management for the software development process.
  • Document control: Management of procedures, work instructions, templates, and records produced during development.
  • Supplier controls: Evaluation, qualification, and monitoring of external development partners under the manufacturer's quality system. A contracted development company may be treated as a supplier under the manufacturer's ISO 13485 quality system.
  • Verification and validation: Evidence that software meets requirements and that those requirements are appropriate for the intended use.
  • Corrective and preventive actions: Processes for identifying, investigating, and resolving quality issues.
  • Change management: Controlled processes for managing changes to software requirements, design, code, and release configuration.
  • Quality records: Documented evidence of development activities, review outcomes, test results, and release decisions.

An important distinction for buyers outsourcing development: the medical device manufacturer retains its quality management system responsibilities regardless of which development activities are performed by an external partner. A contracted development company may operate within the manufacturer's quality system as a supplier, contribute to the manufacturer's design and development records, or maintain its own quality management system aligned with the manufacturer's requirements. The allocation of responsibilities should be documented in the development agreement and clearly understood before work begins.

The FDA's Quality Management System Regulation (QMSR) became effective February 2, 2026, and aligns with ISO 13485:2016 more closely than the previous Quality System Regulation (QSR). Manufacturers who have been operating under the previous QSR and development partners working with US-regulated medical device manufacturers should understand the QMSR's requirements and how they affect design and development controls, supplier management, and documentation requirements.

How Does ISO 14971 Support Medical Device Software Risk Management?

ISO 14971:2019 specifies the requirements for a risk management process for medical devices. For software-containing medical devices and SaMD, risk management is not a final documentation step. It runs throughout the software lifecycle and influences architecture decisions, requirements, testing priorities, and release criteria.

The ISO 14971 risk management process includes:

  • Hazard identification: Identifying what could go wrong with the device or software, including software failures, incorrect outputs, missing outputs, data integrity issues, and unsafe interactions with other systems.
  • Risk estimation: Estimating the probability of harm and the severity of that harm for each identified hazard situation.
  • Risk evaluation: Determining whether estimated risks are acceptable within the risk policy defined in the risk management plan.
  • Risk control: Implementing measures to reduce risks to acceptable levels, which may include software design decisions, architectural choices, alert mechanisms, user interface constraints, or operational restrictions.
  • Verification of risk controls: Confirming that risk controls have been implemented correctly and that they are effective.
  • Residual risk: Evaluating remaining risk after controls are applied and making a documented determination that residual risk is acceptable.
  • Production and postproduction information: Monitoring the device in use for new hazards, defects, and changing risk profiles, and feeding that information back into the risk management process.

For software, risk management should influence architecture from the outset. A diagnostic software function whose failure could affect a clinical decision requires different architectural safeguards, testing rigor, and release criteria than a utility function with no direct patient impact. A development partner who treats risk management as a documentation exercise performed after development is complete is not applying risk management in the way ISO 14971 or IEC 62304 intend.

How Should Medical Device Software Be Verified, Validated, and Tested?

Verification and validation are distinct activities in the medical device software context, and the distinction matters.

Software verification is the confirmation, through objective evidence, that the outputs of a software development phase meet the requirements for that phase. It answers the question: did we build the software correctly? Verification activities include unit testing, integration testing, code review, and static analysis.

Software validation is the confirmation, through objective evidence, that software that is to be used as or in a medical device fulfills its intended use and user needs. It answers the question: did we build the right software? Validation connects the software to its intended medical purpose and clinical context.

These are different from clinical validation or clinical evaluation, which address whether the device performs its medical purpose safely and effectively in the intended clinical population.

Testing activities for medical device software may include:

  • Requirements-based testing: Test cases derived from documented software requirements, ensuring that every requirement is tested.
  • Unit testing: Testing of individual software units in isolation. Required at Class B and C under IEC 62304, with additional rigor at Class C.
  • Integration testing: Testing of software components combined into the integrated system, verifying that interfaces and interactions between components work as specified.
  • System testing: Testing of the complete, integrated software system against all software requirements.
  • Regression testing: Confirming that software changes have not introduced new failures in previously tested functionality.
  • Risk-based testing: Allocating additional testing effort to software functions associated with higher-risk hazard scenarios.
  • Usability testing: Evaluating the user interface against usability engineering requirements, where applicable under IEC 62366-1.
  • Security testing: Testing for vulnerabilities, penetration resistance, and correct implementation of security controls.
  • Performance and reliability testing: Confirming that the software performs correctly under expected load conditions and handles failure scenarios appropriately.

Test traceability connects requirements, identified risks, test cases, and test results. A traceability matrix shows that every requirement has been tested, that risk controls have been verified, and that the testing evidence supports the release decision. This traceability is typically part of the software documentation package relevant to a premarket submission.

Hypothetical example: A diagnostic support software function analyzes a patient's lab results against reference ranges and generates an alert for a clinician when a value suggests a potential condition. The requirement that the alert fires when a lab value exceeds the threshold is a testable requirement. The risk that the alert fails to fire when it should, possibly delaying clinical intervention, is a software-related hazard that should be identified in the risk analysis. A test case that specifically tests the alert trigger at boundary values, documents the expected result, and records the actual result is the evidence that connects that requirement and that risk to the testing process. If the software is modified to change the alert threshold, regression testing should confirm that the alert behavior remains correct.

What Additional Requirements Apply to AI-Powered SaMD?

AI-enabled medical device software introduces additional considerations that go beyond the standard IEC 62304 lifecycle processes. FDA has been developing its regulatory framework for AI in medical devices over several years, and buyers developing AI-powered SaMD should understand the current state of that framework.

FDA's January 2025 draft guidance on AI-enabled device software functions described an approach to regulating the marketing, labeling, transparency, and performance monitoring of AI-enabled devices. FDA's August 2025 final guidance on Predetermined Change Control Plans (PCCPs) provided a pathway for manufacturers to describe anticipated modifications to AI models in advance, allowing changes within the PCCP scope to proceed without a new premarket submission. Buyers should check FDA's current guidance documents for any subsequent updates, as this area continues to develop.

Key considerations for AI-powered SaMD development:

  • Model development and training: The model development process, including training data selection, preprocessing, feature engineering, and architecture decisions, is part of the design and development record.
  • Training and validation data: The composition, source, representativeness, and quality of training data affect model performance and the generalizability of results.
  • Data representativeness: A model trained on data from specific patient populations, clinical settings, or imaging equipment may perform differently when deployed in a broader or different clinical context.
  • Bias and performance considerations: Model performance should be evaluated across relevant demographic and clinical subgroups. Disparities in performance across subgroups can represent a patient safety concern.
  • Model evaluation: Evaluation metrics should be appropriate for the clinical task. Accuracy alone is rarely sufficient for clinical applications. Sensitivity, specificity, positive and negative predictive value, and their clinical implications should be understood.
  • Explainability: Some clinical applications benefit from, or may require, an ability to explain the basis for a model output. This is a design consideration that affects architecture and documentation.
  • Human oversight: FDA has expressed interest in ensuring that clinicians can independently review and disregard AI outputs. The degree to which the system supports or constrains human review is part of the product design.
  • Model updates and change management: AI models in clinical use may require updates as new data becomes available or as performance drifts. Change management processes must address when a model update constitutes a device modification requiring a new or supplemental submission.
  • Postmarket monitoring: AI-enabled devices that are deployed in clinical settings should have processes for monitoring real-world performance and for detecting and responding to performance degradation or unexpected behaviors.

AI regulation is product-specific: Not all AI-powered medical software follows the same regulatory pathway. Whether an AI-enabled software function is a medical device, which classification applies, and whether a PCCP is relevant are product-specific determinations that require regulatory analysis. FDA's draft guidance documents represent recommendations and proposed approaches; they should not be treated as binding requirements until finalized.

What Cybersecurity Requirements Should Medical Device Software Developers Understand?

Cybersecurity is a patient safety consideration for medical devices, not only an IT security matter. A cybersecurity failure that affects a medical device's performance, data integrity, or availability can result in harm to patients. FDA's approach to medical device cybersecurity reflects this understanding.

FDA's February 2026 Cybersecurity in Medical Devices guidance updates the previous 2023 guidance and addresses cybersecurity considerations for applicable cyber devices, as defined in Section 524B of the FD&C Act. For applicable cyber devices, FDA expects sponsors to submit a plan to monitor, identify, and address postmarket cybersecurity vulnerabilities, a software bill of materials (SBOM), and evidence that the device meets cybersecurity requirements. These are premarket submission requirements for applicable devices, not voluntary recommendations.

For medical device software development, cybersecurity should be addressed throughout the lifecycle, not addressed as a final testing step:

  • Secure architecture: Design decisions should address how the software handles data, communicates with external systems, authenticates users, and enforces access controls.
  • Threat modeling: Identifying potential threats, attack surfaces, and adversarial scenarios relevant to the device's clinical environment and connectivity profile.
  • Authentication and access control: Implementing appropriate mechanisms to authenticate users and systems, and to restrict access to authorized functions and data.
  • Encryption: Protecting data in transit and at rest using appropriate cryptographic methods.
  • Software bill of materials (SBOM): Documenting all software components, including open-source and third-party libraries, their versions, and known vulnerabilities. The SBOM is a postmarket maintenance tool as well as a submission requirement for applicable devices.
  • Vulnerability management: Processes for monitoring for new vulnerabilities in the software and its dependencies, assessing their impact on the device, and responding appropriately.
  • Security testing: Testing that verifies the implementation of security controls and identifies exploitable vulnerabilities.
  • Patch and update processes: A defined process for delivering security patches and software updates that maintains device integrity and meets applicable regulatory requirements for device modifications.
  • Secure development practices: Coding standards, code review practices, and development environment controls that reduce the introduction of vulnerabilities during development.

The specific cybersecurity requirements that apply depend on whether the device is an applicable cyber device under Section 524B, its connectivity characteristics, the clinical environment in which it is used, and the applicable regulatory pathway. A development partner should understand these distinctions and design security controls that are appropriate for the specific product.

How Can Buyers Find a Developer Who Understands Both FDA and IEC 62304?

A developer capable of building regulated medical device software is not simply a healthcare software developer who is familiar with HIPAA. Medical device software development requires a different set of engineering practices, documentation habits, process controls, and regulatory knowledge. Buyers should evaluate the following capabilities before selecting a development partner.

1. Experience with regulated medical device software

Ask whether the company has built SaMD, embedded medical device software, or other regulated medical device software that has reached the market. Relevant experience means working through the full lifecycle, including development, verification, testing, documentation, and release, not only prototype-level work.

2. Familiarity with IEC 62304 lifecycle processes

Ask how the team manages requirements, architectural design, unit implementation, verification, integration testing, system testing, and software maintenance. A team that understands IEC 62304 should be able to describe their process for each lifecycle activity and the documentation they produce.

3. Understanding of FDA expectations

Ask whether the team understands device software classification, intended use, and applicable documentation requirements for the submission type relevant to the buyer's product. The team should understand the difference between guidance and regulation, and between verification and validation.

4. Quality management system experience

Ask how the development team works within an ISO 13485-aligned quality environment. Specifically, ask how they handle design inputs and outputs, review records, change controls, and documentation in a way that supports the manufacturer's quality management system requirements.

5. Risk management capability

Ask how the team collaborates with the manufacturer's quality and regulatory functions on ISO 14971 risk management. Risk management should influence architecture, requirements, and testing decisions, not be completed as a document review after development is finished.

6. Verification and validation evidence

Request examples of traceability matrices, test plans, verification reports, and software validation documentation from past projects, with confidential information appropriately redacted. A team experienced in medical device software development will have these artifacts and should be able to explain how they connect to the development process.

7. Medical device cybersecurity experience

Ask how security risks are identified, documented, tested, and managed throughout development and maintenance. A team that addresses cybersecurity only at the testing phase is not applying it as a design consideration throughout the lifecycle.

8. Change and configuration management

Ask how software versions, component dependencies, changes, defect resolutions, and release decisions are documented and controlled. Configuration management is a core IEC 62304 process and a quality management system requirement. The answer should describe a systematic, documented approach, not an informal practice.

Buyer caution: A development company that claims to be "FDA compliant" or "IEC 62304 certified" without being able to describe its specific processes, documentation outputs, and evidence of past regulated projects is not demonstrating the capability those claims suggest. Request process documentation and redacted project examples rather than accepting general claims of compliance.

What Questions Should Buyers Ask Before Hiring a Medical Device Software Development Company?

1. Have you developed SaMD or software integrated into a regulated medical device?
Credible answer: Describes specific types of products, the development lifecycle activities involved, and can provide redacted examples of documentation or reference past product submissions with appropriate confidentiality protections.
Concern: Claims experience in "healthcare software" or "HIPAA-compliant applications" without describing specific medical device development work.
2. Which IEC 62304 lifecycle processes do you follow?
Credible answer: Describes the planning, requirements, design, implementation, verification, integration testing, system testing, release, and maintenance processes they follow, and can describe the records they produce for each.
Concern: References IEC 62304 by name but cannot describe how specific lifecycle activities are conducted or what documentation they produce.
3. How do you determine and document software safety classification?
Credible answer: Describes a risk-based analysis process that considers the device system, software failure modes, potential harm, and applicable risk controls, and produces documented rationale for the classification of each software item.
Concern: Assigns safety class without documented analysis, or uses simplified rules that do not account for the device context and risk management.
4. How do you manage software requirements and traceability?
Credible answer: Describes a requirements management approach that links requirements to design, implementation, risk controls, and test cases, and produces a traceability matrix that supports verification and release decisions.
Concern: Describes requirements as user stories in a backlog with no formal traceability to testing or risk management outputs.
5. How do you integrate ISO 14971 risk management into development?
Credible answer: Describes risk management as a process that runs throughout development, with software hazard identification, risk estimation, risk control selection, and verification of risk controls integrated into the lifecycle activities.
Concern: Describes risk management as a documentation deliverable produced at the end of a development phase.
6. Have you worked within an ISO 13485 quality management system?
Credible answer: Describes how their development processes interface with a manufacturer's quality management system, including design controls, document control, change management, and record keeping.
Concern: Conflates ISO 13485 with general quality practices or describes no formal quality management system interface.
7. How do you support FDA software documentation requirements?
Credible answer: Describes the documentation their lifecycle processes produce and how that documentation supports a premarket submission, referencing relevant FDA guidance on software documentation.
Concern: Describes "FDA-compliant" development without being able to describe what documentation is produced or how it connects to submission requirements.
8. How do you handle verification, validation, and regression testing?
Credible answer: Distinguishes verification from validation, describes requirements-based test cases, documents test protocols and results, produces a traceability matrix, and describes a regression testing approach for software changes.
Concern: Uses verification and validation interchangeably, or describes testing only as functional QA without connection to requirements, risks, or traceability.
9. How do you manage third-party and off-the-shelf software components?
Credible answer: Describes how third-party components are evaluated, documented, and monitored for defects and vulnerabilities, consistent with FDA's OTS Software guidance and IEC 62304 requirements for software items of unknown provenance.
Concern: Treats open-source or commercial components as outside the scope of their development process without systematic evaluation or documentation.
10. Who is responsible for regulatory strategy and final submission decisions?
Credible answer: Clearly acknowledges that the medical device manufacturer retains regulatory strategy decisions and final submission responsibility. Describes how their development outputs support the manufacturer's regulatory affairs function without overstepping into the manufacturer's responsibilities.
Concern: Claims to handle FDA submission preparation without clarifying that the manufacturer retains regulatory responsibility, or is unclear about where their responsibilities end.

What Should a Medical Device Software Development Proposal Include?

A proposal that promises "FDA-compliant software development" without defining processes, deliverables, and responsibilities does not give a buyer the information needed to evaluate whether the developer understands what they are committing to. A credible proposal should address the following elements:

  • Product description and intended medical use, including the basis for the team's understanding of regulatory scope
  • Regulatory assumptions and dependencies, with acknowledgment of what is outside the developer's scope
  • Technical discovery findings from an assessment of requirements, infrastructure, and third-party dependencies
  • Proposed software architecture with component descriptions
  • IEC 62304 lifecycle approach describing which processes will be followed and what documentation will be produced
  • Software safety classification assumptions with documented rationale
  • Risk management responsibilities and how they interface with the manufacturer's ISO 14971 process
  • Quality management system interfaces, including how the developer's work connects to the manufacturer's ISO 13485 design controls
  • Requirements management and traceability approach
  • Verification and validation plan with defined testing activities and acceptance criteria
  • Cybersecurity approach including threat modeling, security controls, and SBOM
  • Third-party and off-the-shelf software identification and documentation approach
  • Documentation deliverables list, specifying what records the buyer will receive
  • Project milestones with defined inputs, outputs, and acceptance criteria
  • Change management and configuration management approach
  • Maintenance and postmarket support scope
  • IP ownership and documentation handover at project end

What Are the Most Common Questions About SaMD and IEC 62304?

What is Software as a Medical Device?
Software as a Medical Device is software intended for one or more medical purposes that performs those purposes without being part of a hardware medical device. The IMDRF definition, referenced by FDA, focuses on intended medical purpose rather than the platform or setting in which software is used.
Does all medical software require FDA approval?
No. Not all software used in healthcare qualifies as a medical device under FDA's framework. Software that does qualify as a device is subject to risk-based regulation, and many software-containing devices do not require premarket approval (PMA). Whether a specific product requires a premarket submission, and which pathway applies, depends on its device classification, intended use, and applicable exemptions.
What is the difference between SaMD and SiMD?
SaMD is standalone software that performs a medical purpose on a general-purpose computing platform, without being part of a hardware medical device. SiMD is software that is part of a hardware medical device or is necessary for the hardware device to work as intended. Both are subject to medical device regulatory requirements, but they may have different risk profiles and development responsibilities.
What does IEC 62304 require?
IEC 62304 specifies lifecycle processes for medical device software, covering development planning, requirements analysis, architectural design, detailed design, implementation and verification, integration testing, system testing, release, maintenance, risk management, configuration management, and problem resolution. The specific activities required for each process depend on the software safety class assigned to the software item.
Is IEC 62304 mandatory for FDA-regulated software?
IEC 62304 is a recognized consensus standard, not a statutory requirement. FDA recognizes it in its consensus standards database, meaning manufacturers can declare conformity in premarket submissions to satisfy applicable documentation expectations. Following IEC 62304 is not legally required, but it represents an accepted approach to producing medical device software documentation and development controls that align with FDA's expectations.
What are IEC 62304 Classes A, B, and C?
IEC 62304 defines three software safety classes based on the potential consequences of software failure. Class A applies where no injury or harm is possible from software failure. Class B applies where failure could result in injury but not serious injury. Class C applies where failure could result in death or serious injury. The required development and verification activities increase with class.
Does IEC 62304 cover software validation?
IEC 62304 covers software verification and system testing activities. Software validation, in the FDA context, is a broader activity that connects the software to its intended medical use and user needs. IEC 62304 lifecycle processes contribute to software validation, but validation in the FDA sense also involves connecting the software requirements and testing evidence to the product's intended medical purpose.
What is the difference between IEC 62304 and ISO 13485?
IEC 62304 specifies software lifecycle processes for medical device software. ISO 13485 specifies quality management system requirements for medical device manufacturers and supply chain organizations. IEC 62304 processes are implemented within the quality management system framework that ISO 13485 describes. The two standards are complementary, not interchangeable.
What is the role of ISO 14971 in SaMD development?
ISO 14971 provides the risk management framework for medical devices, including SaMD. IEC 62304 requires that software risk management activities be conducted in coordination with the broader device risk management process described in ISO 14971. Risk management should influence software architecture, requirements, testing, and release decisions throughout the lifecycle.
Can a general software development company build SaMD?
A general software development company may have the technical skills to build a functional software product, but medical device software development requires specific lifecycle processes, documentation practices, quality management system experience, risk management knowledge, and regulatory understanding that general software development does not require. Buyers should evaluate specific medical device software capabilities, not general software quality.
Does AI-powered medical software require FDA clearance?
Not all AI-powered medical software requires FDA clearance. Whether an AI-enabled software function is a regulated medical device, and which regulatory pathway applies, depends on its intended use, its functions, and applicable regulatory criteria. Product-specific regulatory analysis is required. FDA's regulatory framework for AI-enabled devices continues to develop through guidance documents and policy updates.
What should I look for when hiring a medical device software development company?
Look for demonstrated experience with regulated medical device software projects, familiarity with IEC 62304 lifecycle processes, understanding of FDA documentation requirements, quality management system experience, ISO 14971 risk management capability, and medical device cybersecurity practices. Request process documentation and redacted project examples. A credible vendor can describe its processes specifically, not just claim general compliance.

What Is the Final Advice for Companies Building Software as a Medical Device?

Regulated medical software requires more than competent software engineering. It requires a development partner that understands the product's intended medical purpose, can reason about its regulatory classification, follows a documented lifecycle process aligned with IEC 62304, integrates risk management throughout development rather than at the documentation phase, produces evidence of verification and validation connected to requirements and risks, manages cybersecurity as a design consideration, and plans for ongoing maintenance and change management after the software is released.

The medical device manufacturer retains applicable regulatory responsibilities regardless of which development activities are outsourced. That includes regulatory strategy, quality management system oversight, supplier management, and responsibility for the accuracy and completeness of any premarket submission. A development partner can contribute expertise and deliverables, but it cannot take on the manufacturer's regulatory responsibilities. Buyers should understand that division before engaging any external development team.

The practical implication for outsourcing decisions: a development company's capability claims should be evaluated against specific process descriptions, documentation examples, and evidence of past regulated projects, not against general claims of FDA compliance or healthcare software experience. The questions in this article provide a starting point for that evaluation.

Evaluate Medical Device Software Development Partners

TechRadiant's directory of medical device software development companies provides a resource for evaluating development partners with regulated medical device software experience before shortlisting and due diligence.